Measurement of the time resolution of small SiPM-based scintillation counters

E. A. Kravchenko, V. V. Porosev, G. A. Savinov

Результат исследования: Научные публикации в периодических изданияхстатья

1 Цитирования (Scopus)

Аннотация

In this research, we evaluated the timing resolution of SiPM-based scintillation detector on a 1-GeV electron beam "extracted" from VEPP-4M. We tested small scintillation crystals of pure CsI, YAP, LYSO, and LFS-3 with HAMAMATSU S10362-33-025C and S13360-3050CS. The CsI scintillator together with HAMAMATSU S13360-3050CS demonstrated the best results. Nevertheless, the achieved time resolution of ∼80 ps (RMS) relates mainly to the photodetector itself. It makes the silicon photomultiplier an attractive candidate to replace other devices in applications where sub-nanosecond accuracy is required.

Язык оригиналаанглийский
Номер статьи12020
Число страниц14
ЖурналJournal of Instrumentation
Том12
Номер выпуска12
DOI
СостояниеОпубликовано - 14 дек 2017