@article{e6d4919bcc0345dab4afd397012f70dd,
title = "Measurement of the photoelectron collection efficiency in MCP PMT",
abstract = "A method of direct measurement of the photoelectron collection efficiency in microchannel plate photomultiplier tubes (MCP PMT) has been developed. The method was verified by testing a PMT with conventional dynode system. The measurement accuracy of 5% was achieved. Large number of MCP PMTs produced in Novosibirsk (Russia) was tested.",
keywords = "Electron multipliers (vacuum), Photon detectors for UV, visible and IR photons (vacuum), Photon detectors for UV, visible and IR photons (vacuum) (photomultipliers, HPDs, others), visible and IR photons (vacuum) (photomultipliers, HPDs, others), MICROCHANNEL PLATE PHOTOTUBES, Photon detectors for UV",
author = "Barnyakov, {A. Yu} and Barnyakov, {M. Yu} and Cygankov, {D. A.} and Katcin, {A. A.} and Kononov, {S. A.} and Onuchin, {A. P.}",
note = "Publisher Copyright: {\textcopyright} 2017 IOP Publishing Ltd and Sissa Medialab srl. Copyright: Copyright 2018 Elsevier B.V., All rights reserved.",
year = "2017",
month = mar,
day = "27",
doi = "10.1088/1748-0221/12/03/P03027",
language = "English",
volume = "12",
journal = "Journal of Instrumentation",
issn = "1748-0221",
publisher = "IOP Publishing Ltd.",
number = "3",
}