Аннотация
The new spectrometric channel for the Spherical Neutral Detector (SND) electromagnetic calorimeter (EMC) is described. It provides measurement of the calorimeter signal arrival time and amplitude with 1 ns and 250 keV resolutions, respectively. The new electronics is useful for reliable detection of low-speed anti-neutrons from the e +e −→nn̄ reaction near threshold and needed for increasing the EMC operation rate. The algorithm of determination of signal parameters (time, amplitude) is based on invariability of the signal waveform. We present the procedure of the waveform calibration using signals from cosmic-muon and Bhabha scattering events.