A Picosecond Image-Converter Dissector with a Crossed Sweep

S. V. Andreev, N. S. Vorobiev, V. L. Dorokhov, A. I. Zarovsky, A. S. Komelkov, G. Ya Kurkin, O. I. Meshkov, A. V. Smirnov

Результат исследования: Научные публикации в периодических изданияхстатья

Аннотация

A dual-slit dissector operating in the crossed-sweep mode with a picosecond time resolution has been created on the basis of the PIF-01 image converter tube. The dissector is applicable to synchrotron radiation sources and electron-positron colliders when it is necessary to simultaneously record longitudinal profiles of multi-bunch beams that fill two to several hundred adjacent separatrices of an accelerator ring. Two trains of light pulses shifted in time with respect to each other have been recorded with a time resolution of 6.0 ± 0.5 ps during the dynamic tests of the developed dissector with a laser system.

Язык оригиналаанглийский
Страницы (с-по)87-91
Число страниц5
ЖурналInstruments and Experimental Techniques
Том63
Номер выпуска1
DOI
СостояниеОпубликовано - янв 2020

Fingerprint Подробные сведения о темах исследования «A Picosecond Image-Converter Dissector with a Crossed Sweep». Вместе они формируют уникальный семантический отпечаток (fingerprint).

  • Цитировать

    Andreev, S. V., Vorobiev, N. S., Dorokhov, V. L., Zarovsky, A. I., Komelkov, A. S., Kurkin, G. Y., Meshkov, O. I., & Smirnov, A. V. (2020). A Picosecond Image-Converter Dissector with a Crossed Sweep. Instruments and Experimental Techniques, 63(1), 87-91. https://doi.org/10.1134/S0020441220010017