Radiation hardness study of CsI(Tl) scintillation crystals for the Belle II calorimeter

I. Chakin, M. Golkovksy, A. Kuzmin, D. Matvienko, E. Sedov, B. Shwartz

Research output: Contribution to journalArticlepeer-review

1 Citation (Scopus)

Abstract

The electromagnetic calorimeter of the Belle II detector contains CsI(Tl) crystals of 30 cm length which have been used at the Belle experiment. We measure the light output degradation of CsI(Tl) crystals exposed to uniformly distributed absorbed dose. Four Belle typical crystals with known scintillation characteristics are irradiated with photons at a total dose of about 35 krad. Results show acceptable radiation hardness for the Belle II experiment conditions where the accumulated dose in crystals could reach 10 krad.

Original languageEnglish
Article number06034
Number of pages7
JournalJournal of Instrumentation
Volume12
Issue number6
DOIs
Publication statusPublished - 26 Jun 2017

Keywords

  • Calorimeters
  • Radiation damage to detector materials (solid state)
  • Radiation-hard detectors

Fingerprint

Dive into the research topics of 'Radiation hardness study of CsI(Tl) scintillation crystals for the Belle II calorimeter'. Together they form a unique fingerprint.