MOCVD of Pt coatings with high surface areas on the contacts of electrophysiological diagnostic electrodes

Research output: Contribution to journalConference articlepeer-review

Abstract

The potential of a chemical vapor deposition technique to produce Pt coatings with high surface area contacts of an electrophysiological diagnostic electrode has been firstly demonstrated. On the first step, a Pd layer has been deposited in the hydrogen atmosphere from Pd(hfac)2 on the contacts made of stainless steel. On the second step, the top functional Pt coatings have been deposited in the oxygen atmosphere from Pt(acac)2. The effect of deposition parameters and the Pd sublayer on the microstructural and electrochemical characteristics of Pt coatings has been studied. The capacity and impedance values of samples are 22-198 mC·cm-2 (0.1M H2SO4, 100 mV/s) and 56-167 Ω (at 120 Hz), respectively.

Original languageEnglish
Pages (from-to)924-930
Number of pages7
JournalMaterials Today: Proceedings
Volume19
DOIs
Publication statusPublished - 1 Jan 2019
Event11th International Conference on Materials Science and Engineering (BraMat) - Brasov, Romania
Duration: 13 Mar 201916 Mar 2019

Keywords

  • Pt coatings
  • Pd sublayer
  • contact of an EFI electrode
  • CV
  • PALLADIUM LAYERS
  • CAPACITANCE
  • PLATINUM
  • Contact of an EFI electrode

Cite this